Strany 1 - RANGE OPTION 0 - 1000 NTU

MICRO 200 BW TURBIDIMETERRANGE OPTION 0 - 1000 NTUHF scientific, inc.3170 Metro ParkwayFt. Myers, FL 33916-7597Phone: (239) 337-2116Fax: (239) 332-7

Strany 3 - Field Service Kit

MICRO 200 BW 0 - 1000 NTU (5/05)Rev. 3.1MICRO 200 BWTABLE OF CONTENTSI. IMPORTANT INFORMATION ...

Strany 5

MICRO 200 BW 0 - 1000 NTU (5/05)Rev. 3.1MICRO 200 BWTABLE OF CONTENTS CONT.IV. ROUTINE OPERATION ...

Strany 7 - DECLARATION OF CONFORMITY

Page 1MICRO 200 BW 0 - 1000 NTU (5/05)Rev. 3.1I. IMPORTANT INFORMATIONA. HOW TO USE THIS MANUALThe equipment has been designed for simple and easy op

Strany 8

Page 2MICRO 200 BW 0 - 1000 NTU (5/05)Rev. 3.1C. SPECIFICATIONS FOR THE MICRO 200 BWSpecification MICRO 200 BWRanges: 0 - 1000.0 NTUMethod: EPA approv

Strany 9

Page 3MICRO 200 BW 0 - 1000 NTU (5/05)Rev. 3.1C. SPECIFICATIONS FOR THE MICRO 200 BW CONTINUEDSpecifications Micro 200 BWShipping Weight: Approximate

Strany 10

Page 4MICRO 200 BW 0 - 1000 NTU (5/05)Rev. 3.1E. FIGURESFigure No. Title Page No.1 Desiccant Tray Installation ...

Strany 11 - TABLE OF CONTENTS

Page 5MICRO 200 BW 0 - 1000 NTU (5/05)Rev. 3.1F. COMPLIANCE TESTING CERTIFICATIONThe Micro 200 BW has undergone extensive testing,therefore, HF scien

Strany 13 - TABLE OF CONTENTS CONT

Page 6MICRO 200 BW 0 - 1000 NTU (5/05)Rev. 3.1II. INSTALLATIONA. PACKING LIST FOR THE MICRO 200 BWCONTENTS QUANTITYInstruction Manual 1MICRO 200 BW 1

Strany 14

Page 7MICRO 200 BW 0 - 1000 NTU (5/05)Rev. 3.1Figure 3Turbidity Sensor Outline DimensionsFigure 2Analyzer Outline Dimensions8.780XX.XX=INCHES(XX)=mm(

Strany 15 - A. HOW TO USE THIS MANUAL

Page 8MICRO 200 BW 0 - 1000 NTU (5/05)Rev. 3.1WARNING: Do not restore power until the accesscover has been replaced and secured.When routing the cabl

Strany 16 - Rev. 3.1

Page 9MICRO 200 BW 0 - 1000 NTU (5/05)Rev. 3.1the voltage. The correct fuse ratings are 5 x 20 mm 1Afast acting for 120 VAC and 5 x 20 mm ½A fast act

Strany 17 - D. OPTIONS:

Page 10MICRO 200 BW 0 - 1000 NTU (5/05)Rev. 3.1WARNING: While the chance is slight, there are faultconditions under which a hazardous volt-age could b

Strany 18 - E. FIGURES

Page 11MICRO 200 BW 0 - 1000 NTU (5/05)Rev. 3.1One of the most important features of the MICRO 200BW is the ease of operation. You simply press theF

Strany 19 - Immunity

Page 12MICRO 200 BW 0 - 1000 NTU (5/05)Rev. 3.1 2. Turbidity DisplayThis is the turbidity display. The display shown above is fairly typical. The di

Strany 20 - C. MOUNTING

Page 13MICRO 200 BW 0 - 1000 NTU (5/05)Rev. 3.13. Main MenuThis is the main menu. It is the starting point for all sub-menu selections. Shown below

Strany 21

Page 14MICRO 200 BW 0 - 1000 NTU (5/05)Rev. 3.14. Alarm MenuThis is the ALARM MENU. The left side of the screen shows the selections as described be

Strany 22

Page 15MICRO 200 BW 0 - 1000 NTU (5/05)Rev. 3.15. Display ParametersThis is the Display parameters Menu. The sub-menu selections are described belo

Strany 23

scientific, inc.3170 Metro ParkwayFt. Myers, FL 33916-7597Phone: (239) 337-2116Fax: (239) 332-7643www.hfscientific.comMICRO 200 BW 0 - 1000 NTU (5/0

Strany 24 - A. THEORY & HOW IT WORKS

Page 16MICRO 200 BW 0 - 1000 NTU (5/05)Rev. 3.1b. Set AverageThis is the Set Average Menu. This is where you can make changes to the averaging period

Strany 25

Page 17MICRO 200 BW 0 - 1000 NTU (5/05)Rev. 3.1a. Screen Brightness/HistoryThis is the screen brightness and clear history menu. In this menu you ca

Strany 26

Page 18MICRO 200 BW 0 - 1000 NTU (5/05)Rev. 3.1c. Calibrate 0 - 1000 NTUThis is the turbidity calibration menu.Sensor Calibrate is provided to adjus

Strany 27 - Set up serial port parameters

Page 19MICRO 200 BW 0 - 1000 NTU (5/05)Rev. 3.1d. Comm PortThis is the communications or serial port menu. This menu will have no purpose unless you

Strany 28 - Flashing cursor

Page 20MICRO 200 BW 0 - 1000 NTU (5/05)Rev. 3.1g. Set Access CodeThis screen is encountered after the previous screen when changing the Access Code.

Strany 29

Page 21MICRO 200 BW 0 - 1000 NTU (5/05)Rev. 3.1h. Self Test (4 - 20 mA Cal.)This is the self test menu. Upon entering this menu all internal critic

Strany 30

Page 22MICRO 200 BW 0 - 1000 NTU (5/05)Rev. 3.17. Wait MessagesThese screens indicate that the analyzer is performing some calculations or storing i

Strany 31

Page 23MICRO 200 BW 0 - 1000 NTU (5/05)Rev. 3.18. Backwash DisplayThis is the backwash display. It is only meaningful if Backwash option is connecte

Strany 32

Page 24MICRO 200 BW 0 - 1000 NTU (5/05)Rev. 3.1a. Backwash CalibrationThis is the backwash calibration menu and usable only with Backwash Sensor conn

Strany 33

Page 25MICRO 200 BW 0 - 1000 NTU (5/05)Rev. 3.1B. START UPBefore placing the MICRO 200 BW on line we recom-mend that you standardize the instrument.

Strany 35

Page 26MICRO 200 BW 0 - 1000 NTU (5/05)Rev. 3.1discharge side of the flow through unit using thestainless steel flow control valve (Catalog #50004)sup

Strany 36

Page 27MICRO 200 BW 0 - 1000 NTU (5/05)Rev. 3.1Figure 11Vaporpurge

Strany 37

Page 28MICRO 200 BW 0 - 1000 NTU (5/05)Rev. 3.1is accomplished by washing the interior and exterior ofthe cuvette in a detergent solution, then rinsin

Strany 38

Page 29MICRO 200 BW 0 - 1000 NTU (5/05)Rev. 3.1b. Standard Formazin SolutionsCalibration of this instrument is based on Formazin, amaterial which is

Strany 39 - THROUGH UNIT

Page 30MICRO 200 BW 0 - 1000 NTU (5/05)Rev. 3.1VI. SECURITY ACCESS CODEThe security access code for the MICRO 200 BWprovides a degree of protection ag

Strany 40 - V. ROUTINE MAINTENANCE

Page 31MICRO 200 BW 0 - 1000 NTU (5/05)Rev. 3.1"Sensor Fail"VII. TROUBLESHOOTINGA. GENERAL NOTESService and technical support are availabl

Strany 41 - Vaporpurge

Page 32MICRO 200 BW 0 - 1000 NTU (5/05)Rev. 3.1B. SPARE PARTS LIST FOR MICRO 200 BWCATALOG NO. DESCRIPTION20892 BNC Connector50036 Cuvette (3 pk) 28

Strany 42 - C. CALIBRATION PROCEDURES

Page 33MICRO 200 BW 0 - 1000 NTU (5/05)Rev. 3.1VIII. OPTIONSBACKWASHA. INSTALLATION1. Backwash InterfaceTo monitor backwash using the MICRO 200 BW,

Strany 43

Page 34MICRO 200 BW 0 - 1000 NTU (5/05)Rev. 3.1about three inches. Care must be used to not nick thewires under the jacket. Note that the sensor cable

Strany 44 - VI. SECURITY ACCESS CODE

Page 35MICRO 200 BW 0 - 1000 NTU (5/05)Rev. 3.1These options may be used to connect to a serial printer,a data logger or as a direct interface to a c

Strany 45 - A. GENERAL NOTES

MICRO 200 BW 0 - 1000 NTU (5/05)Rev. 3.1FOREWORDHF TURBIDIMETERSHF turbidimeters are manufactured to meet design criteria for nephelometers as descri

Strany 46

Page 36MICRO 200 BW 0 - 1000 NTU (5/05)Rev. 3.1between the most remote MICRO 200 BW analyzer andthe host computer does not exceed 50 feet (16.4 m.).Du

Strany 47 - A. INSTALLATION

Page 37MICRO 200 BW 0 - 1000 NTU (5/05)Rev. 3.14-21/64(110)3-35/64(90)8-13/16(224)7-3/32(180)Figure 16Backwash Interface Outline DimensionsFigure 17B

Strany 48 - E. BACKWASH REMOTE SWITCHING

Page 38MICRO 200 BW 0 - 1000 NTU (5/05)Rev. 3.13-35/64(90)4-21/64(110)8-13/64(224)7-3/32(180)Figure 18Junction Box OutlineFigure 19Junction Box Wiring

Strany 49 - SERIAL INTERFACES

Page 39MICRO 200 BW 0 - 1000 NTU (5/05)Rev. 3.1SENSOR CABLE TO ANALYZERSENSOR CABLE TO ANALYZERSENSORCHAMBERMOUNTING BRACKETTAKES 1-1/2 IN.PVC PIPESS

Strany 50

Page 40MICRO 200 BW 0 - 1000 NTU (5/05)Rev. 3.1OUTLET BAFFLEINLET BAFFLESENSORMOUNTING KITEXTENSION HANDLEIN KIT)PIPE(NOT SUPPLIEDTO ANALYZERSENSOR CA

Strany 51

Page 41MICRO 200 BW 0 - 1000 NTU (5/05)Rev. 3.1UNITSINTERVENINGLASTVOLTAGERS2321ST OR HUBVOLTAGERS2321845TXDSHIELDGRD RXDRXDGRDDLEIHSTXDTXDGRDRXDCPUH

Strany 52

Page 42MICRO 200 BW 0 - 1000 NTU (5/05)Rev. 3.1WARRANTYHF scientific, inc., as vendor, warrants to the original purchaser of the instruments to be fre

Strany 53 - SENSOR CABLE TO ANALYZER

Page 43MICRO 200 BW 0 - 1000 NTU (5/05)Rev. 3.1JTU: Jackson Turbidity UnitsLCD Liquid Crystal Display (screen) usedDisplay: to view the information

Strany 54

Page 44MICRO 200 BW 0 - 1000 NTU (5/05)Rev. 3.1

Strany 55 - ADDRESS #1

MICRO 200 BW 0 - 1000 NTU (5/05) Rev. 3.1Date: August 3, 2004 Revision: 10 MATERIAL SAFETY DATA SHEET HF scientific, inc. 24 Hour Emergency

Strany 57 - GLOSSARY

MICRO 200 BW 0 - 1000 NTU (5/05)Rev. 3.1DECLARATION OF CONFORMITYApplication of Council Directive: 73/23/EECStandard to which Conformity is Declared

Strany 59

MICRO 200 BW 0 - 1000 NTU (5/05)Rev. 3.1DECLARATION OF CONFORMITYApplication of Council Directive: 73/23/EECStandard to which Conformity is Declared