Hf-scientific Micro200BW Turbidimeter - 0-100 NTU Uživatelský manuál

Procházejte online nebo si stáhněte Uživatelský manuál pro Měřící nástroje Hf-scientific Micro200BW Turbidimeter - 0-100 NTU. HF scientific Micro200BW Turbidimeter - 0-100 NTU User Manual Uživatelská příručka

  • Stažení
  • Přidat do mých příruček
  • Tisk

Shrnutí obsahu

Strany 1 - RANGE 0 - 10 AND 0 - 100 NTU

CATALOG NO. 21652 (05/05)Rev. 3.4MICRO 200 BW TURBIDIMETERRANGE 0 - 10 AND 0 - 100 NTUHF scientific, inc.3170 Metro ParkwayFt. Myers, FL 33916-7597Pho

Strany 2

MICRO 200 BW 0-10 & 0-100 NTU (05/05)Rev. 3.4

Strany 3 - CERTIFICATION

MICRO 200 BW 0-10 & 0-100 NTU (05/05)Rev. 3.4MICRO 200 BWTABLE OF CONTENTS CONT.5.0 ROUTINE MAINTENANCE ...

Strany 4

MICRO 200 BW 0-10 & 0-100 NTU (05/05)Rev. 3.4

Strany 5 - DECLARATION OF CONFORMITY

MICRO 200 BW 0-10 & 0-100 NTU (05/05)Rev. 3.41.0 IMPORTANT INFORMATION1.1 HOW TO USE THIS MANUALThe equipment has been designed for simple and

Strany 6

MICRO 200 BW 0-10 & 0-100 NTU (05/05)Rev. 3.41.3 SPECIFICATIONS FOR THE MICRO 200 BWSpecification MICRO 200 BWRanges: 0 - 10.0000 NTU & 0 -

Strany 7

MICRO 200 BW 0-10 & 0-100 NTU (05/05)Rev. 3.41.3 SPECIFICATIONS FOR THE MICRO 200 BW CONTINUEDSpecifications Micro 200 BWShipping Weight: Appro

Strany 8

MICRO 200 BW 0-10 & 0-100 NTU (05/05)Rev. 3.41.5 FIGURESFigure No. Title Page No.1 Desiccant Tray Installation ...

Strany 9 - TABLE OF CONTENTS

MICRO 200 BW 0-10 & 0-100 NTU (05/05)Rev. 3.41.6 COMPLIANCE TESTING CERTIFICATIONThe Micro 200 BW has undergone extensive testing,therefore, HF

Strany 10 - Rev. 3.4

MICRO 200 BW 0-10 & 0-100 NTU (05/05)Rev. 3.42.0 INSTALLATION2.1 PACKING LIST FOR THE MICRO 200 BWCONTENTS QUANTITYInstruction Manual 1MICRO

Strany 11 - TABLE OF CONTENTS CONT

MICRO 200 BW 0-10 & 0-100 NTU (05/05)Rev. 3.4Figure 3Turbidity Sensor Outline DimensionsFigure 2Analyzer Outline DimensionsPage 78.780XX.XX=INCH

Strany 12

MICRO 200 BW 0-10 & 0-100 NTU (05/05)Rev. 3.4

Strany 13 - 1.1 HOW TO USE THIS MANUAL

MICRO 200 BW 0-10 & 0-100 NTU (05/05)Rev. 3.4Figure 4Installation DiagramPage 8

Strany 14

MICRO 200 BW 0-10 & 0-100 NTU (05/05)Rev. 3.4Figure 5Access Cover RemovalWARNING: Do not restore power until the access coverhas been replaced an

Strany 15 - 1.4 OPTIONS:

MICRO 200 BW 0-10 & 0-100 NTU (05/05)Rev. 3.42.4.2 Analyzer PowerThe power cord (120V) provided is 6 feet (1.83 meters)long. The analyzer power r

Strany 16 - 1.5 FIGURES

MICRO 200 BW 0-10 & 0-100 NTU (05/05)Rev. 3.4Figure 9Remove jumper J-13 to establish isolationof the 4-20mA signal.WARNING: While the chance is s

Strany 17

MICRO 200 BW 0-10 & 0-100 NTU (05/05)Rev. 3.4particles in suspension (turbidity). The optical signalfrom the photodiodes is amplified and then con

Strany 18

MICRO 200 BW 0-10 & 0-100 NTU (05/05)Rev. 3.4Page 133.2.1 Turbidity DisplayThis is the turbidity display. The display shown above is fairly typi

Strany 19

MICRO 200 BW 0-10 & 0-100 NTU (05/05)Rev. 3.43.2.2 Main MenuThis is the main menu. It is the starting point for all sub-menu selections. Shown be

Strany 20

MICRO 200 BW 0-10 & 0-100 NTU (05/05)Rev. 3.4Page 15This is the Sensor Setup menu. As shown above, the cursor is highlighting 10 NTU Turbidity w

Strany 21 - 2.4 ANALYZER CONNECTIONS

MICRO 200 BW 0-10 & 0-100 NTU (05/05)Rev. 3.4By pressing in Normal Cal... menu. - 10 NTU Turbidity OnlyIn this menu you can offset t

Strany 22

MICRO 200 BW 0-10 & 0-100 NTU (05/05)Rev. 3.4Page 17This is the Alarm Setup menu. The left side of the screen shows the selections described bel

Strany 23 - 3.0 STEP-BY-STEP OPERATION

MICRO 200 BW 0-10 & 0-100 NTU (05/05)Rev. 3.4FOREWORDHF TURBIDIMETERSHF turbidimeters are manufactured to meet design criteria for nephelometers

Strany 24

MICRO 200 BW 0-10 & 0-100 NTU (05/05)Rev. 3.4By pressing in Alarm Setup menu. Turbidity/BackwashIn this menu you can set the access

Strany 25 - - - - - - - - - - - - - -

MICRO 200 BW 0-10 & 0-100 NTU (05/05)Rev. 3.43.2.5 Display Parameters - By Pressing from Main MenuThis is the Display Parameters menu. The le

Strany 26

MICRO 200 BW 0-10 & 0-100 NTU (05/05)Rev. 3.4By pressing inside the Display Parameters menu. - Turbidity/BackwashIn this menu you can ch

Strany 27

MICRO 200 BW 0-10 & 0-100 NTU (05/05)Rev. 3.4This is the Monitor Setup menu. The left side of the screen shows the selections described below.

Strany 28

MICRO 200 BW 0-10 & 0-100 NTU (05/05)Rev. 3.43.2.6 Monitor Setup - ContinuedBy pressing in the Monitor Setup menu.In this menu the

Strany 29

MICRO 200 BW 0-10 & 0-100 NTU (05/05)Rev. 3.4By pressing in the Communications menu.In this menu you can select two time periods t

Strany 30

MICRO 200 BW 0-10 & 0-100 NTU (05/05)Rev. 3.4This is the backwash display. It is only meaningful if the Backwash option is connected. The screen a

Strany 31

MICRO 200 BW 0-10 & 0-100 NTU (05/05)Rev. 3.43.3.2 StandardizingThe calibration standard (cuvette) supplied with theMICRO 200 BW is a pure liqui

Strany 32

MICRO 200 BW 0-10 & 0-100 NTU (05/05)Rev. 3.4the desiccant on a quarterly basis or if after cleaning thecuvette readings appear erroneously high t

Strany 33

MICRO 200 BW 0-10 & 0-100 NTU (05/05)Rev. 3.4Figure 12VaporpurgePage 27

Strany 34

MICRO 200 BW 0-10 & 0-100 NTU (05/05)Rev. 3.4

Strany 35

MICRO 200 BW 0-10 & 0-100 NTU (05/05)Rev. 3.4oughly 8 to 10 rinses in clean, distilled water, to removeall streaks.Figure 13Critical Measurement

Strany 36

MICRO 200 BW 0-10 & 0-100 NTU (05/05)Rev. 3.4Page 29It is well known that diluted Formazin is unstable. IfFormazin is used to calibrate the inst

Strany 37 - Calibration Standard Indexing

MICRO 200 BW 0-10 & 0-100 NTU (05/05)Rev. 3.4Page 30MenuMenuMenu8. Press the . The Frequency at the time of thisselection is saved as

Strany 38 - Flow Through Cuvette Indexing

MICRO 200 BW 0-10 & 0-100 NTU (05/05)Rev. 3.4Page 317.0 TROUBLESHOOTING7.1 GENERAL NOTESService and technical support are available from the

Strany 39

MICRO 200 BW 0-10 & 0-100 NTU (05/05)Rev. 3.47.2 SPARE PARTS LIST FOR MICRO 200 BWCATALOG NO. DESCRIPTION20892 BNC Connector50036 Cuvette (3 pk)

Strany 40 - 5.3 CALIBRATION PROCEDURES

MICRO 200 BW 0-10 & 0-100 NTU (05/05)Rev. 3.48.0 OPTIONSBACKWASH8.1 INSTALLATION8.1.1 Backwash Interface (Catalog No. 19275)To monitor backwas

Strany 41 - ones must be made

MICRO 200 BW 0-10 & 0-100 NTU (05/05)Rev. 3.4this point the outer jacket of the cable should be strippedback about three inches. Care must be used

Strany 42 - 6.0 SECURITY ACCESS CODE

MICRO 200 BW 0-10 & 0-100 NTU (05/05)Rev. 3.4contacts are still connected and active unless set to OFFin either of the modes.8.5.1.3 Analog Outp

Strany 43 - 7.1 GENERAL NOTES

MICRO 200 BW 0-10 & 0-100 NTU (05/05)Rev. 3.44-21/64(110)3-35/64(90)8-13/16(224)7-3/32(180)Figure 16Backwash Interface Outline DimensionsFigure 17

Strany 44

MICRO 200 BW 0-10 & 0-100 NTU (05/05)Rev. 3.43-35/64(90)4-21/64(110)8-13/64(224)7-3/32(180)Figure 18Junction Box OutlineFigure 19Junction Box Wir

Strany 45 - Analyzer BNC Connection

MICRO 200 BW 0-10 & 0-100 NTU (05/05)Rev. 3.4DECLARATION OF CONFORMITYApplication of Council Directive: 73/23/EECStandard to which Conformity is

Strany 46 - 8.4 BACKWASH ALARMS

MICRO 200 BW 0-10 & 0-100 NTU (05/05)Rev. 3.4SENSOR CABLE TO ANALYZERSENSOR CABLE TO ANALYZERSENSORCHAMBERMOUNTINGBRACKETTAKES 1-1/2 IN.PVC PIPESS

Strany 47

MICRO 200 BW 0-10 & 0-100 NTU (05/05)Rev. 3.4OUTLET BAFFLEINLET BAFFLESENSORMOUNTING KITEXTENSION HANDLEIN KIT)PIPE(NOT SUPPLIEDTO ANALYZERSENSOR

Strany 48

MICRO 200 BW 0-10 & 0-100 NTU (05/05)Rev. 3.48.0 OPTIONS - ContinuedSERIAL INTERFACES8.6 SERIAL INTERFACESEach of these options maybe installed

Strany 49

MICRO 200 BW 0-10 & 0-100 NTU (05/05)Rev. 3.4to the RS-232 Standard Voltage Mode Interfaceexcept that it transmits in a “Balanced” mode, in onedi

Strany 50

MICRO 200 BW 0-10 & 0-100 NTU (05/05)Rev. 3.4Page 42UNITSINTERVENINGLASTVOLTAGERS2321ST OR HUBVOLTAGERS2321845TXDSHIELDGRD RXDRXDGRDDLEIHSTXDTXDGR

Strany 51

MICRO 200 BW 0-10 & 0-100 NTU (05/05)Rev. 3.4WARRANTYHF scientific, inc., as vendor, warrants to the original purchaser of the instruments to be

Strany 52 - 8.6 SERIAL INTERFACES

MICRO 200 BW 0-10 & 0-100 NTU (05/05)Rev. 3.4LCD Liquid Crystal Display (screen) used toDisplay view the information available on theMICRO 200 BW

Strany 53

Control # 0501 Date: August 3, 2004 Revision: 10 MATERIAL SAFETY DATA SHEET HF scientific, inc. 24 Hour Emergency 3170 Metro Parkway

Strany 54

MICRO 200 BW 0-10 & 0-100 NTU (05/05)Rev. 3.4

Strany 55 - WARRANTY

MICRO 200 BW 0-10 & 0-100 NTU (05/05)Rev. 3.4DECLARATION OF CONFORMITYApplication of Council Directive: 73/23/EECStandard to which Conformity is

Strany 56 - GLOSSARY

MICRO 200 BW 0-10 & 0-100 NTU (05/05)Rev. 3.4

Strany 57

MICRO 200 BW 0-10 & 0-100 NTU (05/05)Rev. 3.4MICRO 200 BWTABLE OF CONTENTS1.0 IMPORTANT INFORMATION .........

Komentáře k této Příručce

Žádné komentáře