OWNER’S MANUAL HF – Micro 100 Laboratory Turbidimeter HF scientific 3170 Metro Parkway Fort Myers, Florida 33916-7597
Micro 100 (5/10) Page 4 Rev. 3.3 Figure 2 illustrates all the items that can appear on the display. The upper row of the display (1) is used fo
Micro 100 (5/10) Page 5 Rev. 3.3 7. If you are measuring and comparing more than one sample, pressing the enter key will display the latest readi
Micro 100 (5/10) Page 6 Rev. 3.3 If the Micro 100 will be used over the entire range of .02 to 1000 NTU a complete calibration as described below wi
Micro 100 (5/10) Page 7 Rev. 3.3 2. The turbidity value displayed in the lower row of the display should read 1000 NTU. This is the first
Micro 100 (5/10) Page 8 Rev. 3.3 6.0 User Selectable Parameters The MICRO 100 provides you the ability to customize your instrument according to y
Micro 100 (5/10) Page 9 Rev. 3.3 6.4 Setting the Calibration Interval After pressing the key, the upper row of the display will have the letters
Micro 100 (5/10) Page 10 Rev. 3.3 7.0 Troubleshooting 7.1 System Warning Message(s) Automatic warning messages are generated by the MICRO 100 to
Micro 100 (5/10) Page 11 Rev. 3.3 If errors 1-4 are noted, turn the instrument off and then back on. Once the instrument is back on, re-calibrate
Micro 100 (5/10) Page 12 Rev. 3.3 Lamp Replacement cont’d. 1. Remove the lamp module from the instrument by squeezing the two side tabs on the
Micro 100 (5/10) Page 13 Rev. 3.3 9.0 Contacting the HF scientific Technical Service Department For technical assistance or to order replacement
Micro 100 (5/10) Page 14 Rev. 3.3 Glossary Formazin Turbidity Units (FTU): see Nephelometric Turbidity Units Indexing a Cuvette: The United St
Micro 100 (5/10) Page 15 Rev. 3.3 WARRANTY HF scientific, as vendor, warrants to the original purchaser of the instruments to be free of defec
Micro 100 (5/10) Rev 3.3 DECLARATION OF CONFORMITY Application of Council Directive: 73/23/EEC Standard to which Conformity is Declared: Prod
Micro 100 (5/10) Rev. 3.3 TABLE OF CONTENTS SPECIFICATIONS...
Micro 100 (5/10) Rev. 3.3 8.0 ROUTINE MAINTENANCE ... 11 8.
Micro 100 (5/10) Page 1 Rev. 3.3 SPECIFICATIONS: MICRO 100 Conforms to specifications set forth in EPA method 180.1 (Nephelometric Method) § MI
Micro 100 (5/10) Page 2 Rev. 3.3 1.0 Using This Instruction Manual Congratulations on your purchase of a new MICRO 100 or MICRO 100 IR Laboratory
Micro 100 (5/10) Page 3 Rev. 3.3 3.0 Becoming Familiar with the Instrument Figure 1 is a depiction of the front of the MICRO 100. Not shown in th