8. Pour the 10.0 NTU calibration standard into the sample well (ensure that the well is
completely filled). Place the black cleanout plug back on the sample well (to block
light).
Note: If you choose to use Formazin, be sure to clean the sample well immediately after
using Formazin and prior to returning to the measurement mode.
9. Press the
button to indicate to the instrument that you wish to calibrate on the 10.0
NTU calibration level. Once you have pressed the
button, the current reading will
be actively displayed on the LCD display.
Wait for the reading to stabilize, once the reading has stabilized press the
button
again. The word CAL will appear on the upper row of the LCD display and the word
NO will appear on the lower row of the LCD display.
10. This is the final step for calibration at the 10.0 NTU level; if you wish to exit without
calibrating simply press the
button. However, if you are satisfied and wish to
complete calibration on the 10.0 NTU level use the tu arrows to select the word
YES on the lower row of the display and then press the
button to initiate
calibration.
Both the CAL and STORE blocks will flash during the operation and the word YES
will change to a count down starting at 90.